Can I edit or modify factory autosequence from Ansur?
Can I use all the test elements of Ansur ProSim Plug-In inside a step?
i. Any one test element from the ECG Simulation Test group
ii. Respiration Simulation Test
iii. Temperature Simulation Test
iv. Any one of IBP Dynamic Simulation or IBP Static pressure test from the IBP Simulation Test
group
&nnbsp;v. Only NIBP Dynamic Simulation test from the NIBP Test group
vi. SpO2 test
b) If you include test elements other than the above, Ansur displays a message box and discards those test elements
Can we use Ansur Firmware upload to update ProSim 6/8 firmware through wireless?
Firmware update to ProSim 8 can be done ONLY through USB cable.
How can I execute factory autosequences from Ansur?
You can execute factory autosequences in 2 ways
a) 1st way: Open a factory template from the Ansur template library. Click play button from Ansur toolbar and click Start button from the testguide
b) 2nd way: Open Ansur Mini Plug-In. Click Autosequence tab. Select any autosequence from the list. Click Play button. Click start button from the test guide
How many test records can be saved in the ProSim 8 Vital Signs Simulator?
How to create autosequence from Ansur similar to the autosequence present in the <br>ProSim simulator?
To create autosequence similar to the autosequence present in the ProSim
simulator follow the steps below
a) Drag and drop Autosequence container to the template
b) Drag and drop one more autosequence container under it. This becomes the step which simulates the test elements in parallel
c) Drag and drop test elements under the step
d) Save it as custom templates in the Ansur test library or to any other location in PC
How to customize autosequences and presets?<strong></strong></a>
a) End users are able to customize autosequences and presets by using Ansur ProSim 8 Mini Plug-In which comes along with Ansur executive version 2.9.6. Default autosequences and presets are available in Ansur ProSim library after installing Ansur ProSim 6/8 Plug-In.
b) Pre-sets values can be replaced and restored to default values by using ProSim 8 Ansur Mini Plug-in.
c) End users are able to make their customized autosequences by using ProSim 8 Ansur Mini Plug-in and save them in the ProSim 8. The default autosequences cannot be replaced.
In Ansur ProSim Plug-In, what is the total duration of the execution of the step?
a) The total duration of the execution of the step is the sum of the duration set in each test elements under the step
b) If any one of the duration is set as indefinite then the step execution will be continuous until Next button or Stop button is clicked.
What is the difference between execution of tests in serial and parallel in Ansur ProSim <br>Plug-In?
Execution of tests in serial allows the user to simulate waveforms individually, and ProSim device resets to LOCAL mode at completion of each test
b) Execution of the tests in parallel allows user to simulate tests simultaneously, and ProSim device resets to LOCAL mode only at the completion of each Step
When to use autosequence container and step in Ansur ProSim Plug-In?
a) If you want to execute tests sequentially one by one and enter the results individually and navigate to next test element manually then, drag and drop the test elements into the template without autosequence container
b) If you want to execute tests sequentially one by one and enter the results individually and navigate to next test element automatically then, drag and drop the test elements into the template under the autosequence container
c) If you want to execute tests in parallel at the same time and enter the results after each step then, drag and drop the appropriate test elements under the step
Where can I find the ProSim 8 Ansur mini plug-in instruction on how to edit and create auto sequences?
- Download both latest Ansur Executive and ProSim 8 Ansur plug-in software versions fromwww.flukebiomedical.comwebsite to your PC.
- From your PC, select START: ALL PROGRAMS, then select the FLUKE tab.
- Both the Ansur Executive and ProSim 8 Ansur plug-in user manual can be found under ANSUR HELP.
- The mini plug-in can be used to create and edit auto sequences, presets and r-curves and upgrade future new firmware. Detailed instruction is in the ProSim 8 Ansur plug-in user manual.
Why am I not able to enter or edit heart rate or test duration values in the edit box in the ProSim 8 Ansur Plug-in?
Why am I not able to get files from ProSim 8 using Ansur ProSim 8 Mini Plug-In?
Do ProSim 4 and ProSim 8 use the same battery charger?
- Made clear mention of it in the user manuals.
- Added a colored flag to the battery charger (ProSim 4 = yellow label, ProSim 8 = red label).
- Designed electrical protection from possible burn-out from application of the wrong battery charger. Please keep in mind: though this protection is in place, using the wrong battery charger can cause loss of some battery life if connected to the wrong voltage for extended periods of time.
For optimum battery life, should the battery pack be kept fully charged?
For optimum battery pack life, should the battery pack be charged after every use, or only charged when the charge level drops below some threshold?
For optimum battery pack life, should the battery pack be disconnected from the charger after the LED turns green?
How does the battery level affect the ProSim performance?
The ProSim power circuits were designed to operate over a specific range of battery voltage levels. As the battery charge is depleted, the battery voltage drops. At a point just above the minimum design voltage for the power supplies, the ProSim will shut itself off. The firmware monitors the battery charge level and warns the customer in advance of this shutdown so that they can respond appropriately (e.g., save date, plug in the charger, swap out the battery pack, etc.).
As long as there is sufficient battery voltage for the ProSim to remain powered on, all circuits will operate normally, even at 0 % charge level. This is because the ProSim doesn't run directly on the battery voltage, because the battery voltage varies and precision circuits need a stable power supply voltage to work properly. The ProSim first converts the variable battery voltage into a stable voltage that doesn't change even when the battery charge level is very low.
If the battery pack is to be shipped, should it be at a certain charge level?
The battery pack is specified for a "cycle life" of 300 cycles. After 300 charge-discharge cycles, what percentage of the original battery pack capacity will still be usable, typically?
When a battery pack has only 50 % of its original capacity remaining (due to age or the number of charging cycles), what will the gauge read when the battery is fully charged?
Will it harm or shorten the life of the battery pack to leave the charger input connected for an extended period of time?
Can a cuff size affect an NIBP reading on patient monitors?
Improper cuff size has been considered one of the most common errors with blood pressure measurements. It is well documented that readings can be falsely elevated, sometimes to the extreme with a cuff that is too small. The rule established by the American Heart Association (AHA) suggests cuff width should be at least 40 % of the greatest arm circumference.
How are blood pressure readings interpreted?
How does a doctor determine a patient has a normal or abnormal blood pressure?
Doctors have determined a normal range for both systolic and diastolic blood pressure after examining the blood pressure of many people. The following figures can be used as a guide for adult blood pressure:
1. Normal blood pressure: less than 120/80 mmHg
2. High/Normal: between 120/80 and 140/90 mmHg
3. High: equal to or more than 140/90 mmHg
4. Very high: equal to or more than 180/110
Individuals whose blood pressure is consistently higher than this norm are said to have high blood pressure, or hypertension. Uncontrolled high blood pressure is indirectly responsible for many deaths and disability resulting from heart attack, stroke, and kidney failure.
How is human blood pressure generated?
What are the major standards for NIBP monitoring?
The design of the NIBP parameter of a patient monitor is required to be proven to be safe and effective for use in the diagnosis, and treatment of disease states and underlying chronic conditions, and in the prescribing of medications and other therapies. This testing is governed by national and international standards including AAMI/ANSI SP10 and IEC 80601-1-2-30 for minimum performance and safety. The performance of NIBP monitors are required to be evaluated or tested during scheduled maintenance by the manufacturer.
Why do my monitor's BP determinations vary?
Section 3.4.3 of the ANSI Standard for Electronic or Automated Sphygmomanometers specifies the required efficacy of the blood pressure
determination:
"The mean difference of the paired measurements of the test system and the comparison system shall be ± 5 mmHg or less with a standard deviation of8 mmHg or less."
This means variations in individual readings of 5, 6, or even 10 mmHg are normal and do not indicate either the monitor or the Tester are malfunctioning. Some monitors are more repeatable than others, and repeatability is one measure of the overall quality of the monitor. This is why all NIBP standards only emphasize on dynamic pressure repeatability, not on absolute accuracy.
BP Pump2 and CuffLink has a default pulse volume ranging from 0.5 to 0.75 ml, while ProSim is default at 1 ml. Depending on type of monitors, if you change the pulse volume, cuff size (mandrel blocks) or tubing, the absolute NIBP readings may change accordingly, but repeatability should stay the same.
Why do NIBP readings on patient monitors vary from brand to brand?
Each individual manufacturer utilizes their proprietary algorithm for identification of the systolic and diastolic points. Some may use the height of the pressure waveform to mark these points; others may use the slope of the pressure signal. Given that algorithms vary, it is difficult to compare between different blood pressure monitors from different manufacturers.
Why do ProSim NIBP simulated readings not match BP Pump 2 or CuffLink readings on patient monitors?
There are many factors such as tubing, cuff size, and motion, which may affect NIBP readings. This is why all NIBP standards emphasize dynamic pressure repeatability rather than absolute accuracy. BP Pump 2 and CuffLink have a default pulse volume ranging from 0.5 to 0.75ml, while ProSim defaults at 1 ml. Depending on type of monitors, if you change the pulse volume, cuff size (mandrel blocks or 500 ml rigid chamber) or tubing, the absolute NIP readings may change accordingly, but repeatability should stay the same.
Why is an oscillometric NIBP reading higher than an ausculatory NIBP reading?
The auscultatory technique is based on the sounds caused by the blood flow through the artery that is surrounded by the cuff. These sounds are known as Korotkoff (K) sounds. In manual blood pressure measurement, these sounds are detected by a human observer using a stethoscope.
The oscillometric technique does not use K sounds to determine blood pressure. The oscillometric technique monitors the changes/vibrations in cuff pressure caused by the flow of blood through the artery. The monitor inflates the cuff to a pressure that occludes the artery.
Variability between auscultatory and oscillometric blood pressure readings may be due to many factors including human observer's skill and hearing sensitivity. The main reason an oscillometric reading is typically higher is that the human ear can only pick up sound, not vibrations. Vibrations in the artery occur before any sounds have appeared, so by the time the human ear hears the Korotkoff sounds, systolic may have already occurred.
I need a simulator that has four or more IBP (Invasive Blood Pressure) channels; ProSim 8 only has two channels.
The more IBP channels a simulator has, the larger and more expensive the simulator will be. The majority of our fleet of customer design partners reported they do not have to use more than two IBP channels at the same time; those who do need to test three or more IBP channels have chosen to test two IBP channels at a time, then two more, and so on. This preference avoids that extra cost, size and weight, and still yields faster/easier testing in the end.
Alternatively, the new ProSim 3 does have four IBP channels thatyou can test all simultaneously.
I was demonstrating ProSim IBP simulation feature to a customer on a Philips patient monitor, the IBP waveform seemed to be correct, but no numerical display occurred. Is there something wrong with the monitor or ProSim?
- Connect ProSim 8 and HP Viridia monitor with proper IBP cables
- Select IBP on ProSim 8, ZERO both IBP channels
- On HP/Viridia, select MODULE SETUP
- Then select PRESS 1 ABP
- Press ZERO XDUCER
- ZERO IN PROCESS should display on the monitor
- Initiate the ProSim 8 IBP simulation. IBP numbers should display on the HP Viridia monitor
What are the two USB ports on ProSim 8 used for?
What does ACLS stand for?
What is Swan-Ganz Procedure?
Which 4 main parameters does ProSim 4 test?
ProSim 4 simulates following 4 parameters:
1) ECG (including normal and Arrhythmias)
2) Respiration
3) IBP (Invasive Blood Pressures)
4) NIBP (Non-Invasive Blood Pressure)
Which 8 main arameters does ProSim 8 test?
ProSim 8 simulates following 8 parameters:
1) ECG (including normal and Arrhythmias),
2) Respiration
3) IBP (Invasive Blood Pressures)/Cardiac Catheterization
4) Temperature
5) Cardiac Output
6) NIBP (Non-Invasive Blood Pressure)
7) SpO2
8) Masimo Rainbow