General Specifications
| Temperature | Operating | 10 °C to 40 °C (50 °F to 104 °F) |
| Storage | -20 °C to +60 °C (-4 °F to 140 °F) | |
| Humidity | 10 % to 90 % non-condensing | |
| Altitude | 3,000 meters (9,843 ft) | |
| Dimensions (L x W x H) | 14.5 cm x 30.2 cm x 8.6 cm (5.7 in x 11.9 in x 3.4 in) | |
| Display | LCD color display | |
| Communication | USB device upstream port | Mini-B connector for control by a computer |
| USB host controller port | Type A, 5 V output, 0.5 A max load. Connector for keyboard, barcode reader, and printer | |
| Wireless | IEEE 82.15.4 for control by a computer | |
| Power | Lithium-ion rechargeable battery | |
| Battery charger | 100 V to 240 V input, 15 V/2.0 A output. For best performance, the battery charger should be connected to a properly-grounded ac receptacle | |
| Battery life | 9 hours (minimum), 100 NIBP cycles typical | |
| Weight | 1.87 kg (4.2 lb) | |
| Safety standards | IEC/EN 61010-1 3rd Edition; Pollution degree 2 CAT None | |
| Certifications | CE, CSA, C-TICK N10140 , RoHS | |
| Electromagnetic compatibility (EMC) | IEC 61326-1:2012 |
Detailed specifications
| Normal-sinus-rhythm waveform | |
|---|---|
| ECG reference | The ECG amplitudes specified are for Lead II (calibration), from the baseline to the peak of the R wave. All other leads are proportional |
| Normal sinus rhythm | 12-lead configuration with independent outputs referenced to right leg (RL). Output to 10 universal ECG jacks, color-coded to AHA and IEC standards |
| High-level output | 0.5 V/mV ± 5 % of the ECG amplitude setting available on a BNC connector |
| Amplitude | 0.05 mV to 0.5 mV (0.05 mV steps); 0.5 mV to 5.0 mV (0.25 mV steps) Other leads are proportional to Lead II (reference lead) in percentage per: Lead I: 70 Lead II: 100 Lead III: 30 Lead V1: 24 Lead V2: 48 Lead V3: 100 Lead V4: 120 Lead V5: 112 Lead V6: 80 |
| Amplitude accuracy | ± (2 % of setting + 0.05 mV) |
| ECG rate | 10 BPM to 360 BPM in 1 BPM steps |
| Rate accuracy | ± 1 % of setting |
| ECG waveform selection | Adult (80 ms) or pediatric (40 ms) QRS duration |
| ST-segment elevation | Adult mode only. -0.8 mV to +0.8 mV (0.1 mV steps). Additional steps: + 0.05 mV and - 0.05 mV |
| Power-on default | 60 BPM, 1.0 mV, adult QRS and ST-segment elevation of 0 mV |
| Pacemaker waveform | ||
|---|---|---|
| Pacer pulse | Amplitude | 0 (off), ± 2, ± 4, ± 6, ± 8, ± 10, ± 12, ± 14, ± 16, ± 18, ± 20, ± 50, ± 100, ± 200, ± 500, and ± 700 mV for lead II (reference lead) |
| Accuracy |
Reference lead II: ± (5 % setting + 0.2 mV)
|
|
| Pacer pulse width | 0.1 ms, 0.2 ms, 0.5 ms, 1 ms, and 2 ms ± 5 % | |
| Paced arrhythmias | Atrial 80 BPM | |
| Asynchronous 75 BPM | ||
| Demand with frequent sinus beats | ||
| Demand with occasional sinus beats | ||
| Atrio-ventricular sequential | ||
| Noncapture (one time) | ||
| Nonfunction | ||
| Power-on default | Amplitude 5 mV, width 1 ms, atrial waveform |
| Arrhythmia | ||
|---|---|---|
| Baseline NSR | 80 BPM | |
| PVC focus | Left focus, standard timing (except where specified) | |
| Supraventricular arrhythmia | Atrial fibrillation (coarse or fine); atrial flutter; sinus arrhythmia; missed beat (one time); atrial tachycardia; paroxysmal atrial tachcardia; nodal rhythm; and supraventricular tachycardia | |
| Premature arrhythmia | Premature atrial contraction (PAC); premature nodal contraction (PNC); PVC1 left ventricular; PVC1 left ventricular, early; PVC1 left ventricular, R on T; PVC2 right ventricular; PVC2 right ventricular, early; PVC2 right ventricular, R on T; and multifocal PVCs | |
| Ventricular arrhythmia | PVCs 6, 12, or 24 per minute; frequent multifocal PVCs; bigeminy; trigeminy; multiple PVCs (one-time run of 2, 5, or 11 PVCs); monoventricular tachycardia (120 to 300 BPM in 5 BPM steps); poly-ventricular tachycardia (5 types); ventricular fibrillation (coarse or fine); and asystole | |
| Conduction defect | First-, second-, or third-degree heart block; and right- or left-bundlebranch block | |
| Advanced cardiac life support | Shockable pulseless arrest rhythms | Ventricular fibrillation (coarse), ventricular fibrillation (fine), unstable polymorphic ventricular tachycardia |
| Non-shockable pulseless arrest rhythms | Asystole | |
| Symptomatic bradycardia | Sinus bradycardia (< 60 BPM) | |
| 2nd degree AV block, mobitz type I | ||
| 2nd degree AV block, mobitz type II | ||
| Complete/3rd degree AV block | ||
| Right bundle branch block | ||
| Left bundle branch block | ||
| Advanced cardiac life support cont. | Symptomatic tachycardia: regular narrow-complex tachycardia (QRS < 0.12 seconds) | Sinus tachycardia > 150 BPM |
| Supraventricular Tachycardia | ||
| Symptomatic tachycardia: regular wide-complex tachycardias (QRS ≥ 0.12 seconds) | Sinus tachycardia > 150 BPM | |
| Supraventricular tachycardia SVT with aberrancy | ||
| Irregular tachycardia | Atrial fibrillation (coarse and fine), atrial flutter, unstable monomorphic ventricular tachycardia (120 BPM to 300 BPM), torsade de pointes/polymorphic ventricular tachycardia (long QT interval) |
| ECG Performance testing | ||
|---|---|---|
| Amplitude |
0.05 mV to 0.5 mV (0.05 mV steps); 0.5 mV to 5.0 mV (0.25 mV steps) Other leads are proportional to Lead II (reference lead) in percentage per: Lead I: 70 |
|
| Pulse wave | 30 BPM, 60 BPM, with 60 ms pulse width | |
| Square wave | 0.125 Hz, 2 Hz, 2.5 Hz | |
| Triangle wave | 0.125 Hz, 2 Hz, 2.5 Hz | |
| Sine wave | 0.05 Hz, 0.5 Hz, 1, 2 Hz, 5 Hz, 10 Hz, 25 Hz, 30 Hz, 40 Hz, 50 Hz, 60 Hz, 100 Hz, and 150 Hz | |
| R-wave detection | Waveform | Triangular pulse |
| Rate | 30 BPM, 60 BPM, 80 BPM, 120 BPM, 200 BPM, and 250 BPM | |
| Width | 8 ms to 20 ms in 2 ms steps, and 20 ms to 200 ms in 10 ms steps | |
| Width accuracy | ± (1 % of setting + 1 ms) | |
| QRS detection | Widths | 8 ms to 20 ms in 2 ms steps, and 20 ms to 200 ms in 10 ms steps |
| Width accuracy | ± (1 % of setting + 1 ms) | |
| Rate | 30 BPM, 60 BPM, 80 BPM, 120 BPM, 200 BPM, and 250 BPM | |
| R-Wave up slope | 0.875 amplitude, 0.4375 x width | |
| R-Wave down slope | Full amplitude, 0.5 x width | |
| S-Wave up slope | 0.125 amplitude, 0.0625 x width | |
| Tall T-wave rejection | Waveform | QT Interval 350 ms |
| T-Wave width 180 ms | ||
| T-Wave shape ½ sinewave | ||
| Amplitude | 0 % to 150 % reference lead amplitude in 10 % steps | |
| Rate | 80 BPM | |
| Rate accuracy | ± 1 % of setting | |
| Amplitute accuracy | ± (2 % of setting + 0.05 mV) |
| ECG Artifact | |
|---|---|
| Type | 50 Hz, 60 Hz, muscular, baseline wander, respiration |
| Size | 25 %, 50 %, 100 % of the normal sinus R-Wave for each lead |
| Lead Select | All, RA, LL, LA, V1, V2, V3, V4, V5, V6 |
| Fetal/Maternal ECG | |
|---|---|
| Fetal Heart Rate (fixed) | 60 BPM to 240 BPM in 1 BPM steps |
| Fetal Heart Rate (IUP) | 140 BPM at beginning, then varies with pressure |
| Intrauterine-Pressure Waveforms | Variable deceleration, early deceleration, late deceleration, and uniform acceleration |
| Wave Duration | 90 seconds, bell-shaped pressure curve, from 0 mmHg to 90 mmHg and returning to 0 |
| IUP Period | 2 min, 3 min, or 5 minutes; and manual |
| Default Settings | FHR 120 BPM, uniform deceleration wave, manual |
| Invasive Blood Pressure | |
|---|---|
| Channels | 2, each independently settable with identical parameters and are individually electrically isolated from all other signals |
| Input/Output Impedance | 300 Ω - or ± 10 % |
| Exciter Input Range | 2 to 16 V peak |
| Exciter-Input Frequency Range | DC to 5000 Hz |
| Transducer Sensitivity | 5 (default) or 40 µV/V/mmHg |
| Pressure Accuracy | ± (1 % of setting + 1 mmHg) accuracy guaranteed for dc excitation only |
| Static Pressure | - 10 to + 300 mmHg in 1 mmHg steps |
| Pressure Units | mmHg or kPa |
| Dynamic Waveforms | Types (default pressures) |
|
| Pressure variability | Systolic and diastolic pressures are independently variable in 1 mmHg steps |
| Swan-Ganz Sequence | Right atrium, right ventrical (RV), pulmonary artery (PA), pulmonary artery wedge (PAW) |
| Cardiac Catheterization | Chambers |
Aortic, pulmonary valve, and mitral valve |
| Respiration Artifact | Arterial, radial artery, and left ventricle |
5 % to 10 % multiplication |
| Other | 5 mmHg or 10 mmHg |
| BP Output | Circular DIN 5-Pin |
| Power-On Default | 0 mmHg |
| Respiration | |
|---|---|
| Rate | 0 (OFF), 10 BrPM to 150 BrPM in 1 BrPM steps |
| Waves | Normal or ventilated |
| Ratio (inspiration:expiration) | Normal |
1:1, 1:2, 1:3, 1:4, 1:5 |
| Ventilated |
1:1 |
| Impedance Variations (? Ω) | 0.00 Ω to 1.00 Ω iin 0.05 Ω steps and 1Ω to 5 Ω in 0.25 Ω steps |
| Accuracy Delta | ± (3 % of setting + 0.05 Ω) |
| Baseline | 500 Ω, 1000 Ω (default), 1500 Ω, 2000 Ω, Leads I, II, III |
| Accuracy Baseline | ±5 % |
| Respiration Lead | LA or LL (default) |
| Apnea Selection | 12 sec, 22 sec, or 32 seconds (one-time events), or continuous (Apnea ON = respiration OFF) |
| Power-On Default | 20 BrPM, delta 1.0 Ω |
| Temperature | |
|---|---|
| Temperature | 30 °C to 42.0 °C in 0.5 °C steps |
| Accuracy | ± 0.4 °C |
| Compatibility | Yellow Springs, Inc. (YSI) Series 400 and 700 |
| Output | Circular DIN 4-Pin |
| Cardiac Output | |
|---|---|
| Catheter Type | Baxter Edwards, 93a-131-7f |
| Calibration Coeffecient | 0.542 (0 °C injectate), 0.595 (24 °C injectate) |
| Blood Temperature | 36 °C (98.6 °F) to 38 °C (100.4 °F) ± 2 % in 1 °C steps |
| Injectate Volume | 10 cc |
| Injectate Temperature | 0 °C or 24 °C |
| Cardiac Output | 2.5, 5, 10 liters per minute ± 7.5 % |
| Faulty-Injectate Curve | Waveform for simulation available |
| Left-To-Right-Shunt Curve | Waveform for simulation available |
| Calibrated Pulse | 1.5 ° for 1 second |
| Connector | Circular DIN 7 pin |
| Power-On Default | 5 liters per minute, 0 °C injectate, 37 °C blood temperature |
| Non-Invasive Blood Pressure | ||
|---|---|---|
| Pressure Units |
mmHg or kPa |
|
| Pressure Relief Test Range | 100 to 400 mmHg | |
| NIBP Simulations | Pulse | 2 mmHg max into 500 ml NIBP system |
| Volume of air moved | 1.25 ml max | |
| Simulations (systolic/diastolic [MAP]) |
|
|
| Repeatability | Within ± 2 mmHg (at maximal pulse size independent of device under test) | |
| Synchronization: normal Sinus heart rates: 30 BPM to 240 BPM |
|
|
| Synchronization: Arrhythmias | Premature atrial contraction (PAC), premature ventricular contraction (PVC), atrial fibrillation, and missed beat | |
| Leak Test | Target pressure | 20 to 400 mmHg |
| Elapse time | 0:30 to 5:00 minutes: seconds in 30 second steps | |
| Leakage rate | 0 mmHg/minute to 200 mmHg/minute | |
| Rate | 30 BPM, 60 BPM, 80 BPM, 120 BPM, 200 BPM, and 250 BPM | |
| R-Wave up slope | 0.875 amplitude, 0.4375 x width | |
| R-Wave down slope | Full amplitude, 0.5 x width | |
| S-Wave up slope | 0.125 amplitude, 0.0625 x width | |
| Manometer (Pressure Meter) | Range | 10 mmHg to 400 mmHg |
| Resolution | 0.1 mmHg | |
| Accuracy | ± (0.5 % reading + 0.5 mmHg) | |
| Pressure Source | Target pressure range | 20 mmHg to 400 mmHg |
| Resolution | 1 mmHg |
| SpO2Test (Optional) | ||
|---|---|---|
| % O2 |
Range |
30 % to 100 % |
| Resolution | 1 % | |
| % O2Accuracy | With oximeter manufacturer's R-curve |
|
| With Fluke Biomedical R-curves |
|
|
| Heart Rate |
30 BPM to 300 BPM in 1 BPM steps. SpO2 test is synchronized with ECG rate delayed by 150 ms, accuracy ±1% of setting |
|
| Transmission: ratio of detector current to LED current, expressed in parts per million (ppm) | Range | 0 ppm to 300.00 ppm |
| Resolution |
0.01 ppm |
|
| Accuracy | + 50 %/- 30 % for compatible monitors, unspecified for others. Selected by finger size and color: dark, thick finger, medium finger, light, thin finger, neonatal foot. The full range and resolution are available in the engineering mode | |
| Pulse Amplitude | Range | 0 % to 20.00 % |
| Resolution | 0.01 % | |
| Artifact | Respiration |
|
| Ambient light |
|
|
| Masimo Rainbow Technology | Test Masimo Rainbow technology with an optional adapter supplied by Masimo that allows the ProSim two wavelength to test the Rainbow multiple wavelength system | |
| Compatible Manufacturer Products | With manufacturer R-curve | Nellcor, Masimo, Nonin, and Nihon Kohden |
| With Fluke R-curve | Mindray, GE-Ohmeda, Philips/HP, and BCI |
Pre-Defined Simulations
- Normal
- Hypertensive
- Hypotensive
- Tachycardic
- Bradycardic
- Ventricular Fibrillation
- Asystole
Autosequences
- Monitor Testing Sequence
- ECG Sequence
- Oximeter Testing Sequence
- Cardiac Failure Sequence
- Arrhythmia Sequence
- Exercise Sequence
- Respiration Sequence
- NIBP Testing Sequence
- IBP Testing Sequence
- Temperature Sequence
Next Article: ProSim8 Resources